Additional Service Procedures
9G9002 – 2011-08-15
6-11
Checking Image Quality
[1]
Ask the customer to send images from the modalities to the laser imager.
[2]
Check for image skew. If necessary, adjust the optics assembly (see
“Imaging Assembly—Start Index Delay
Adjustment”
).
[3]
Is the image quality acceptable to the customer?
[4]
Click the
Image Optimization
tab on the SCP Configuration screen.
[5]
At the top of the screen, select an SCU from the drop-down list.
Note
• The information for each entry in the SCU drop-down list includes, in order: IP Address and AE Title of the SCU
that will send the image, and AE Title of the destination.
• The parameter values on the screen apply to the modality identified in the SCU list.
• Perform
Image Optimization
and
Other Options
for each SCU in the drop-down list.
[6]
Click
Modify
.
[7]
To change a parameter, type the new value or select it from the drop-down list.
Note
• The value selected will be the default for this SCU.
• The value in the list to the right of the parameter value indicates whether the value sent from the SCU is
overridden.
– If you select
Always Apply
, the parameter on the screen overrides the parameter the SCU sends.
– If you select
No
, the SCU parameter is applied. Normally, you should select
No
to allow the SCU to override
the default selected.
Yes
No
Advance to
“Editing the Service
History Log”
.
Continue with Step
4
.