Diagnostics
9G9002 – 2011-08-15
4-65
Condition Code: 27-607 – Optics Maximum Laser Power Outside Valid Range
Condition Code: 27-608 – Invalid Laser Dynamic Range
Web Portal Message
Log Message
DICOM Printer
StatusInfo
Severity
Optics: Calibration
Failed
Optics calibration
detected Optics
Maximum Power
outside valid range
ELEC DOWN
0-Fatal
Description
This error occurs when the maximum laser power measured by the optics
calibration and diagnostics is outside the predefined valid range.
The error clears when a calibration is requested or power is cycled.
System Impact
The imager can only print jobs that were in process when the error occurred.
The spooler can still accept jobs, but the jobs will not print until the error is
resolved. All jobs that were in process when the error occurred will reprint
completely when the error is resolved. [SI-1]
User Action
1. Shut down the laser imager, and then restart the imager.
2. If the error persists, call for service.
Show at DS
True
DICOM Printer
Status
FAILURE
Unique Symb. ID
ERR_OPTICS_MAX_LASER_POWER
Logged for RMS
[TriggerValues=x][TriggerRange=y,z][TriggerNotes=Laser Power: value, limits]
Web Portal Message
Log Message
DICOM Printer
StatusInfo
Severity
None
Optics calibration
detected laser dynamic
range approaching
predefined limits
NORMAL
3-Notify
Description
This notification occurs when the laser dynamic range has been adjusted and
its value is within 10% of the upper or lower limit of the adjustment range.
System Impact
None. The system will function normally. [SI-0]
User Action
None
FE Action
None. This is a log notification only.
Show at DS
False
DICOM Printer
Status
NORMAL
Unique Symb. ID
ERR_OPTICS_LASER_DYNAMIC_RANGE_NEAR_LIMIT