SERVICE MANUAL
2-40
9G9002 – 2011-08-15
Optics and Exposure Transport
Removal
[1]
Remove 2 screws on each side of the imaging assembly where the optics and exposure transport are joined.
[2]
Carefully separate the optics from the transport.
Installation
Reverse the steps in the removal procedure.
Note
Use the alignment pins near the screw holes to help line up the two components.
Prerequisites:
Remove the imaging assembly
.
Tools:
T20 Torx driver
Screws
Screws
Exposure Transport
Optics