SERVICE MANUAL
2-2
9G9002 – 2011-08-15
To Check:
[1]
Energize the laser imager and let it warm to operating temperature.
[2]
Prepare the meter:
(a)
Install the block on the probe 140.0 mm (5.5 in.) from the probe sensor.
(b)
Clean the probe with alcohol wipes.
(c)
Set the meter to display in °C.
[3]
Remove the back cover.
Block
Probe Sensor
Probe
Meter
140.0 mm (5.5 in.)