Appendix D
145
Error Messages
Error Messages
D.
Er
ro
r M
e
ssa
g
e
s
-161
Invalid block data
Block data are expected, but the block data received are invalid for some reason. (Refer to
IEEE488.2,7.7.6.2.) For example, the END message is received before the length of the
block data is reached.
-171
Invalid expression
The equation data element is invalid. (Refer to IEEE488.2,7.7.7.2.) For example,
parentheses are not paired or a character violates the standard.
-213
Init ignored
Another measurement has been being executed and the measurement start request
(
:INITiate[:IMMediate]
command ) has been ignored.
-224
Illegal parameter value
The parameter recieved is not correct. For example, though a correct program message was
“
:CALC1:FORM CP
,” a wrong program message, “
:CALC1:FORM RP
,” was received.
L
46
LOAD measurement incomplete
This error occurs when the cable correction LOAD measurement is incomplete.
M
43
Measurement failed
A measurement failure has occurred during measuring the correction data.
-109
Missing parameter
The number of parameters is less than required by the command. For example, although
the
:CREJ:LIM
command requires one parameter such as “
:CREJ:LIM 3
”, no parameter
is added to give “
:CREJ:LIM
”.
-250
Mass storage error
An error occurred while accessing the external mass storage device.
N
83
No data to load
There is no setup data for the selected number or no external mass storage device is
connected.
-120
Numeric data error
Numeric data is improper.
-128
Numeric data not allowed
A numeric value data element (that does not violate the standard) has been received where
the E4981A does not accept any numeric value data element.
Summary of Contents for E4981A
Page 1: ...Keysight E4981A 120 Hz 1 kHz 1 MHz Capacitance Meter Service Guide ...
Page 2: ......
Page 13: ...Contents 11 L 145 M 145 N 145 O 146 P 146 Q 146 R 147 S 147 T 147 U 148 Warning Message 149 ...
Page 14: ...12 Contents ...
Page 70: ...68 Chapter4 Troubleshooting Performance test failure troubleshooting ...
Page 84: ...82 Chapter5 Replaceable Parts Replaceable Parts List Front Panel Figure 5 12 Front Panel ...
Page 132: ...130 Chapter7 Post Repair Procedures Post Repair Procedures ...
Page 138: ...136 AppendixB Firmware Update Update the E4981A firmware using USB cable ...
Page 142: ...140 AppendixC Power Requirement Power Requirements ...