Chapter 2
19
Performance Test
Individual Test Description
3. Ch
apter
T
itle
4. Chapter
T
itle
5. Chapter T
itle
2. Pe
rfor
ma
nce T
e
st
Individual Test Description
1. Test Signal Frequency Accuracy Test
Description
This test checks the frequency accuracy of the E4981A test signal.
Specification
@23
±
5
⋅
C
Test equipment
Test point and limits
Cable Length : 2 m
Frequency [Hz]
Spec.
120 to 1M
*1
±
0.02%
*1.Option 001 only.
Frequency Counter
Keysight 53131A or Keysight 53181A
Interface Box
Keysight 04284-65007
Test Leads (BNC), 2m
Keysight 16048D
BNC(f)-BNC(f) Adapter
Keysight 1250-1830, 4ea
BNC(m)-BNC(m) Cable (122cm)
Keysight 8120-1840
Frequency
Frequency Shift
Test Limit
Measurement
Uncertainty
TAR
120 Hz
0 %
±
0.024 Hz
±
0.00093 Hz
25.88
1 kHz
0 %
±
0.20 Hz
±
0.00784 Hz
25.51
1 MHz
*1
-2 %
±
196 Hz
±
7.93 Hz
24.74
-1 %
±
198 Hz
±
7.93 Hz
24.97
0 %
±
200 Hz
±
7.93 Hz
25.24
+1 %
±
202 Hz
±
7.93 Hz
25.49
+2 %
±
204 Hz
±
7.92 Hz
25.77
*1.Option 001 only.
Summary of Contents for E4981A
Page 1: ...Keysight E4981A 120 Hz 1 kHz 1 MHz Capacitance Meter Service Guide ...
Page 2: ......
Page 13: ...Contents 11 L 145 M 145 N 145 O 146 P 146 Q 146 R 147 S 147 T 147 U 148 Warning Message 149 ...
Page 14: ...12 Contents ...
Page 70: ...68 Chapter4 Troubleshooting Performance test failure troubleshooting ...
Page 84: ...82 Chapter5 Replaceable Parts Replaceable Parts List Front Panel Figure 5 12 Front Panel ...
Page 132: ...130 Chapter7 Post Repair Procedures Post Repair Procedures ...
Page 138: ...136 AppendixB Firmware Update Update the E4981A firmware using USB cable ...
Page 142: ...140 AppendixC Power Requirement Power Requirements ...