Required Adjustment after Replacing Assembly
After replacing the analog assembly (PPMC, A31 and A34), all adjustment must be
required. The adjustment must be performed Keysight’s qualified service personnel. If you
need the adjustment for your E4981A, it should be sent to the nearest Keysight
Technologies service office.
Writing ID
This item writes the serial number, option information into the A31 board of the E4981A.
Required equipment for the writing ID
None
Test Signal Level Monitor Adjustment
This adjustment calibrate the Signal Level Monitor’s absolute measurement accuracy.
Required equipment for the Signal Level Monitor Adjustment
Test Signal Level Adjustment
This adjustment calibrate the Signal Level absolute measurement accuracy.
Required equipment for the Test Signal Level adjustment
Description
Recommended Model
Multimeter
Keysight 3458A
Interface Box
Keysight 04284-65007
BNC(m)-BNC(m) Cable (122cm)
Keysight 8120-1840
Dual Banana-BNC(f) Adapter
Keysight 1251-2277
Description
Recommended Model
Multimeter
Keysight 3458A
Interface Box
Keysight 04284-65007
BNC(m)-BNC(m) Cable (122cm)
Keysight 8120-1840
Dual Banana-BNC(f) Adapter
Keysight 1251-2277
Summary of Contents for E4981A
Page 1: ...Keysight E4981A 120 Hz 1 kHz 1 MHz Capacitance Meter Service Guide ...
Page 2: ......
Page 13: ...Contents 11 L 145 M 145 N 145 O 146 P 146 Q 146 R 147 S 147 T 147 U 148 Warning Message 149 ...
Page 14: ...12 Contents ...
Page 70: ...68 Chapter4 Troubleshooting Performance test failure troubleshooting ...
Page 84: ...82 Chapter5 Replaceable Parts Replaceable Parts List Front Panel Figure 5 12 Front Panel ...
Page 132: ...130 Chapter7 Post Repair Procedures Post Repair Procedures ...
Page 138: ...136 AppendixB Firmware Update Update the E4981A firmware using USB cable ...
Page 142: ...140 AppendixC Power Requirement Power Requirements ...