20
Chapter 2
Performance Test
Individual Test Description
2. Test Signal Level Accuracy Test
Description
This test checks the signal level accuracy of the E4981A source port output signal.
Specification
@23
±
5
⋅
C
Test equipment
Signal Level [V]
Spec.
0.1 to 1.0
±
5.0 %
Digital Multimeter
Keysight 3458A
Interface Box
Keysight 04284-65007
Test Leads (BNC), 2m
Keysight 16048D
BNC(f)-BNC(f) Adapter
Keysight 1250-1830, 4ea
BNC(m)-BNC(m) Cable (122cm)
Keysight 8120-1840
Dual Banana-BNC(f) Adapter
Keysight 1251-2277