92
Chapter 6
Replacement Procedure
3/4 Rack Module Top Replacement
3/4 Rack Module Top Replacement
Tools Required
•
Torque screwdriver, TORX T20
Removal Procedure
,
and for this procedure.
Step 1.
Disconnect the power cable from the E4981A.
Step 2.
Remove the handle, front bumper and rear bumper. (item 1, 2, 3 )
NOTE
To remove the handle, rotate the handle to a vertical position and pull the arms outwards.
Step 3.
Work the six TORX T20 screws (item 4) loose from the bottom.
Step 4.
Remove the 3/4 Rack Module Top.
Replacement Procedure
Step 1.
Reverse the order of the removal procedure.
NOTE
Fasten the 3/4 Rack Module Top from the Rack Module Bottom in the spcified order.
(
Summary of Contents for E4981A
Page 1: ...Keysight E4981A 120 Hz 1 kHz 1 MHz Capacitance Meter Service Guide ...
Page 2: ......
Page 13: ...Contents 11 L 145 M 145 N 145 O 146 P 146 Q 146 R 147 S 147 T 147 U 148 Warning Message 149 ...
Page 14: ...12 Contents ...
Page 70: ...68 Chapter4 Troubleshooting Performance test failure troubleshooting ...
Page 84: ...82 Chapter5 Replaceable Parts Replaceable Parts List Front Panel Figure 5 12 Front Panel ...
Page 132: ...130 Chapter7 Post Repair Procedures Post Repair Procedures ...
Page 138: ...136 AppendixB Firmware Update Update the E4981A firmware using USB cable ...
Page 142: ...140 AppendixC Power Requirement Power Requirements ...