Chapter 4
65
Troubleshooting
Function Specific Troubleshooting
4. T
roub
lesho
o
ting
Step 4.
Replace the Interface Board.
Step 5.
Set DC power supply output v5 V. Connect TP2 (GND) on the scanner interface
simulator to ‘-’ terminal of the power supply. Then connect TP1 (Vcc) on the simulator to
‘+’ terminal of the power supply. (Refer to
.)
Figure 4-12
Scanner Simulator Connections
NOTE
DC power for the scanner interface simulator can be supplied from the E4981A instead of
an external DC power supply.
Step 6.
Connect the scanner simulator to the scanner interface connector on the E4981A’s rear
panel shown in
.
Step 7.
Connect the 1 k
Ω
standard to the UNKNOWN terminal on the E4981A’s front panel.
Step 8.
Turn the E4981A ON.
Step 9.
Press [System].
Step 10.
Press the
SELF TEST
softkey to display the
SELF TEST
page.
Step 11.
Use the
CURSOR
arrow keys to move the cursor to the
TEST MENU
field.
3
On
4
Off
Table 4-10
Setting Bit Switch (S3)
Bit
Default
e4981ase1045
㪛㪚㩷㪧㪦㪮㪜㪩㩷㪪㪬㪧㪧㪣㪰
㪫㫆
㪪㪚㪘㪥㪥㪜㪩
㪠㪥㪫㪜㪩㪝㪘㪚㪜
㪚㫆㫅㫅㪼㪺㫋㫆㫉
㪪㪚㪘㪥㪥㪜㪩㩷㪪㪠㪤㪬㪣㪘㪫㪦㪩
㪜㪋㪐㪏㪈㪘
㪋㪉㪇㪊㪎㪘
䇭㪈䌫㱅
628EE
62)0&
SYNC CLK
In / Out
Trigger
Return
7
8
9
4
5
6
1
2
3
0
.
Preset
Display
Format
Meas
Setup
Recall A
Recall B
Save/
Recall
Capacitance Meter
120Hz/1kHz/1MHz
E4981A
System
Local/
Lock
UNKNOWN
Discharge test device before connecting
r
10V Peak Max Output CAT I
Summary of Contents for E4981A
Page 1: ...Keysight E4981A 120 Hz 1 kHz 1 MHz Capacitance Meter Service Guide ...
Page 2: ......
Page 13: ...Contents 11 L 145 M 145 N 145 O 146 P 146 Q 146 R 147 S 147 T 147 U 148 Warning Message 149 ...
Page 14: ...12 Contents ...
Page 70: ...68 Chapter4 Troubleshooting Performance test failure troubleshooting ...
Page 84: ...82 Chapter5 Replaceable Parts Replaceable Parts List Front Panel Figure 5 12 Front Panel ...
Page 132: ...130 Chapter7 Post Repair Procedures Post Repair Procedures ...
Page 138: ...136 AppendixB Firmware Update Update the E4981A firmware using USB cable ...
Page 142: ...140 AppendixC Power Requirement Power Requirements ...