2. Pe
rfor
ma
nce T
e
st
Test point and limits
Cable Length : 2 m
Frequency
Frequency
Shift
Level
Range
Test Limit
Measurement
Uncertainty
TAR
120 Hz
0 %
0.1 V
10 uF
±
5.0 mV
±
0.28 mV
18.26
0.2 V
±
10.0 mV
±
0.15 mV
66.96
0.3 V
±
15.0 mV
±
0.41 mV
36.83
0.4 V
±
20.0 mV
±
0.38 mV
53.25
0.5 V
±
25.0 mV
±
0.23 mV
108.73
0.6 V
±
30.0 mV
±
0.40 mV
76.52
0.7 V
±
35.0 mV
±
0.40 mV
87.67
0.8 V
±
40.0 mV
±
0.45 mV
89.41
0.9 V
±
45.0 mV
±
0.24 mV
191.54
1.0 V
±
50.0 mV
±
0.26 mV
197.42
0.1 V
1 uF
±
5.0 mV
±
0.24 mV
21.52
0.2 V
±
10.0 mV
±
0.45mV
22.46
0.3 V
±
15.0 mV
±
0.34 mV
44.91
0.4 V
±
20.0 mV
±
0.31 mV
65.84
0.5 V
±
25.0 mV
±
0.44 mV
57.42
0.6 V
±
30.0 mV
±
0.33 mV
92.39
0.7 V
±
35.0 mV
±
0.40 mV
88.02
0.8 V
±
40.0 mV
±
0.21 mV
192.88
0.9 V
±
45.0 mV
±
0.23 mV
196.27
1.0 V
±
50.0 mV
±
0.28 mV
184.35
Summary of Contents for E4981A
Page 1: ...Keysight E4981A 120 Hz 1 kHz 1 MHz Capacitance Meter Service Guide ...
Page 2: ......
Page 13: ...Contents 11 L 145 M 145 N 145 O 146 P 146 Q 146 R 147 S 147 T 147 U 148 Warning Message 149 ...
Page 14: ...12 Contents ...
Page 70: ...68 Chapter4 Troubleshooting Performance test failure troubleshooting ...
Page 84: ...82 Chapter5 Replaceable Parts Replaceable Parts List Front Panel Figure 5 12 Front Panel ...
Page 132: ...130 Chapter7 Post Repair Procedures Post Repair Procedures ...
Page 138: ...136 AppendixB Firmware Update Update the E4981A firmware using USB cable ...
Page 142: ...140 AppendixC Power Requirement Power Requirements ...