32
Chapter 2
Performance Test
Individual Test Description
Cable Length : 2 m, MEAS TIME : 1
Range : 1 uF
Cable Length : 2 m, MEAS TIME : 8
Range : 1 uF
Cable Length : 2 m, MEAS TIME : 8
Range : 1 uF
Cable Length : 2 m, MEAS TIME : 8
Range : 2.2 uF
Standard
Signal
Level
Freq
Freq
Shift
Param
Test Limit
Measurement
Uncertainty
TAR
1 uF
1 V
120 Hz
0 %
Cp
±
0.85 nF
±
70.05 pF
12.13
0 %
D
±
0.00065
±
0.000042
15.79
Standard
Signal
Level
Freq
Freq
Shift
Param
Test Limit
Measurement
Uncertainty
TAR
1 uF
1 V
120 Hz
0 %
Cp
±
0.7 nF
±
69.30 pF
10.10
0 %
D
±
0.0005
±
0.000023
22.55
1 kHz
0 %
Cp
±
0.7 nF
±
44.52 pF
15.72
0 %
D
±
0.0005
±
0.000032
16.02
Standard
Signal
Level
Freq
Freq
Shift
Param
Test Limit
Measurement
Uncertainty
TAR
1 uF
0.1 V
120 Hz
0 %
Cp
±
2.05 nF
±
73.32 pF
27.21
0 %
D
±
0.0019
±
0.000042
44.94
Standard
Signal
Level
Freq
Freq
Shift
Param
Test Limit
Measurement
Uncertainty
TAR
1 uF
1 V
120 Hz
0 %
Cp
±
0.88 nF
±
70.82 pF
12.42
0 %
D
±
0.00068
±
0.000042
16.51
Summary of Contents for E4981A
Page 1: ...Keysight E4981A 120 Hz 1 kHz 1 MHz Capacitance Meter Service Guide ...
Page 2: ......
Page 13: ...Contents 11 L 145 M 145 N 145 O 146 P 146 Q 146 R 147 S 147 T 147 U 148 Warning Message 149 ...
Page 14: ...12 Contents ...
Page 70: ...68 Chapter4 Troubleshooting Performance test failure troubleshooting ...
Page 84: ...82 Chapter5 Replaceable Parts Replaceable Parts List Front Panel Figure 5 12 Front Panel ...
Page 132: ...130 Chapter7 Post Repair Procedures Post Repair Procedures ...
Page 138: ...136 AppendixB Firmware Update Update the E4981A firmware using USB cable ...
Page 142: ...140 AppendixC Power Requirement Power Requirements ...