28
Chapter 2
Performance Test
Individual Test Description
Cable Length : 2 m, MEAS TIME : 1
Range : 100 pF
Cable Length : 2 m, MEAS TIME : 8
Range : 100 pF
Cable Length : 2 m, MEAS TIME : 8
Range : 100 pF
Standard
Signal
Level
Freq
Freq
Shift
Param
Test Limit
Measurement
Uncertainty
TAR
100 pF
1 V
1 MHz
*1
0 %
Cp
±
85 fF
±
11.75 fF
7.23
0 %
D
±
0.00065
±
0.000065
10.03
*1.Option 001 only.
Standard
Signal
Level
Freq
Freq
Shift
Param
Test Limit
Measurement
Uncertainty
TAR
100 pF
1 V
1 kHz
0 %
Cp
±
85 fF
±
8.65 fF
9.82
0 %
D
±
0.00065
±
0.000062
10.61
1 MHz
*1
-2 %
Cp
±
70 fF
±
10.63 fF
6.58
-2 %
D
±
0.0005
±
0.000065
7.71
-1 %
Cp
±
70 fF
±
11.11 fF
6.30
-1 %
D
±
0.0005
±
0.000065
7.71
0 %
Cp
±
70 fF
±
11.15 fF
6.28
0 %
D
±
0.0005
±
0.000065
7.71
1 %
Cp
±
70 fF
±
10.90 fF
6.42
1 %
D
±
0.0005
±
0.000065
7.71
2 %
Cp
±
70 fF
±
10.92 fF
6.41
2 %
D
±
0.0005
±
0.000065 f
7.71
*1.Option 001 only.
Standard
Signal
Level
Freq
Freq
Shift
Param
Test Limit
Measurement
Uncertainty
TAR
100 pF
0.1 V
1 MHz
*1
0 %
Cp
±
205 fF
±
17.79 fF
11.52
0 %
D
±
0.0019
±
0.000065
28.55
*1.Option 001 only.
Summary of Contents for E4981A
Page 1: ...Keysight E4981A 120 Hz 1 kHz 1 MHz Capacitance Meter Service Guide ...
Page 2: ......
Page 13: ...Contents 11 L 145 M 145 N 145 O 146 P 146 Q 146 R 147 S 147 T 147 U 148 Warning Message 149 ...
Page 14: ...12 Contents ...
Page 70: ...68 Chapter4 Troubleshooting Performance test failure troubleshooting ...
Page 84: ...82 Chapter5 Replaceable Parts Replaceable Parts List Front Panel Figure 5 12 Front Panel ...
Page 132: ...130 Chapter7 Post Repair Procedures Post Repair Procedures ...
Page 138: ...136 AppendixB Firmware Update Update the E4981A firmware using USB cable ...
Page 142: ...140 AppendixC Power Requirement Power Requirements ...