Chapter 2
29
Performance Test
Individual Test Description
3. Ch
apter
T
itle
4. Chapter
T
itle
5. Chapter T
itle
2. Pe
rfor
ma
nce T
e
st
Cable Length : 2 m, MEAS TIME : 8
Range : 220 pF
Cable Length : 2 m, MEAS TIME : 8
Range : 470 pF
Standard
Signal
Level
Freq
Freq
Shift
Param
Test Limit
Measurement
Uncertainty
TAR
100 pF
1 V
1 MHz
*1
-2 %
Cp
±
88 fF
±
13.16 fF
6.68
-2 %
D
±
0.00068
±
0.000065
10.49
-1 %
Cp
±
88 fF
±
13.70 fF
6.42
-1 %
D
±
0.00068
±
0.000065
10.49
0 %
Cp
±
88 fF
±
13.64 fF
6.45
0 %
D
±
0.00068
±
0.000065
10.49
1 %
Cp
±
88 fF
±
17.76 fF
4.95
1 %
D
±
0.00068
±
0.000065
10.49
2 %
Cp
±
88 fF
±
15.33 fF
5.74
2 %
D
±
0.00068
±
0.000065
10.49
*1.Option 001 only.
Standard
Signal
Level
Freq
Freq
Shift
Param
Test Limit
Measurement
Uncertainty
TAR
100 pF
1 V
1 MHz
*1
-2 %
Cp
±
126 fF
±
18.70 fF
6.71
-2 %
D
±
0.0011
±
0.000065
16.28
-1 %
Cp
±
126 fF
±
19.47 fF
6.44
-1 %
D
±
0.0011
±
0.000065
16.28
0 %
Cp
±
126 fF
±
16.97 fF
7.39
0 %
D
±
0.0011
±
0.000065
16.28
1 %
Cp
±
126 fF
±
19.30 fF
6.50
1 %
D
±
0.0011
±
0.000065
16.28
2 %
Cp
±
126 fF
±
15.45 fF
8.12
2 %
D
±
0.0011
±
0.000065
16.28
*1.Option 001 only.
Summary of Contents for E4981A
Page 1: ...Keysight E4981A 120 Hz 1 kHz 1 MHz Capacitance Meter Service Guide ...
Page 2: ......
Page 13: ...Contents 11 L 145 M 145 N 145 O 146 P 146 Q 146 R 147 S 147 T 147 U 148 Warning Message 149 ...
Page 14: ...12 Contents ...
Page 70: ...68 Chapter4 Troubleshooting Performance test failure troubleshooting ...
Page 84: ...82 Chapter5 Replaceable Parts Replaceable Parts List Front Panel Figure 5 12 Front Panel ...
Page 132: ...130 Chapter7 Post Repair Procedures Post Repair Procedures ...
Page 138: ...136 AppendixB Firmware Update Update the E4981A firmware using USB cable ...
Page 142: ...140 AppendixC Power Requirement Power Requirements ...