54
Chapter 4
Troubleshooting
Troubleshooting Using Service Function
###: Most suspicious assembly
##: Suspicious assembly
#: Possible faulty assembly
User
1
User Data
User Data Lost
###
2
State Data
State Data Lost
###
3
Correction Data
Corr Data Lost
###
4
Scanner Correction Data
Scanner Data Lost
###
Battery
1
A1_DCBUS_BATTERY
Battery Low
###
Table 4-3
Power on test / self test failure troubleshooting
information
Test
group
Test
No.
Test Name
Error Code /
Message
Probable faulty board
assembly
A31 A34
PPMC