Chapter 4
49
Troubleshooting
Troubleshooting Using Service Function
4. T
roub
lesho
o
ting
Executing the self-test from the front panel
The self-test of the E4981A can be
executed from the front panel. The steps are given below.
Self-test procedure
Step 4.
Press [System] -
SELF TEST
Step 5.
Use the cursor key to highlight the selection item to the right of “TEST No.” (1 in
)
Step 6.
Select the number corresponding to the test that you want to perform by using the
INCR+
and
DECR-
softkeys.
Step 7.
Press
EXECUTE
-
TEST START
(2 in
) to execute the selected test item.
Figure 4-4
Self-test screen and procedure
SERVICE Page
The results of the power on test or the self test (SYSTEM, USER DATA, and BATTERY)
can be saved into a USB memory. Pressing the [System] key followed by the
SERVICE
softkey opens the SERVICE page.
This page is read-only.
This page displays the system information of E4981A and allows you to save the displayed
information into the USB memory, but does NOT allow you to modify or delete the
information.
SCANNER
INTERFACE
Checks the scanner interface. (Visual checks only; no result
shown on screen.)
Table 4-1
Test item
Description
Summary of Contents for E4981A
Page 1: ...Keysight E4981A 120 Hz 1 kHz 1 MHz Capacitance Meter Service Guide ...
Page 2: ......
Page 13: ...Contents 11 L 145 M 145 N 145 O 146 P 146 Q 146 R 147 S 147 T 147 U 148 Warning Message 149 ...
Page 14: ...12 Contents ...
Page 70: ...68 Chapter4 Troubleshooting Performance test failure troubleshooting ...
Page 84: ...82 Chapter5 Replaceable Parts Replaceable Parts List Front Panel Figure 5 12 Front Panel ...
Page 132: ...130 Chapter7 Post Repair Procedures Post Repair Procedures ...
Page 138: ...136 AppendixB Firmware Update Update the E4981A firmware using USB cable ...
Page 142: ...140 AppendixC Power Requirement Power Requirements ...