Chapter 6
115
Replacement Procedure
Power Inlet Assembly Replacement
6. Re
placem
ent Pro
c
e
d
u
re
Power Inlet Assembly Replacement
Tools Required
•
Torque screwdriver, TORX T10
•
Torque screwdriver, TORX T20
•
Cutting plier
Removal Procedure
Refer to
and for this procedure.
Step 1.
Remove the 3/4 rack module top as described in
“3/4 Rack Module Top Replacement” on
Step 2.
Remove the PSU Deck Assembly as described in
“PSU Deck Assembly Removale” on
Step 3.
Cut the two cable ties (item 1).
Step 4.
Remove the two TORX T10 screws (item 2) fastening the Inlet Assembly.
Step 5.
Remove the Inlet Assembly from the Rear Panel.
Step 6.
Detach the fuse box (item 3) from the Inlet Assembly and remove the fuse.
Replacement Procedure
Step 1.
Reverse the order of the removal procedure.
NOTE
Do not forget to insert the fuse in the new Inlet Assembly.