-440
Query UNTERMINATED after indefinite response
In a certain program message, a Query that requests an ambiguous response has not yet
been completely executed when a different Query is received. (Refer to
IEEE488.2,6.5.7.5.7.)
R
16
Reference Measurement Aborted
This error occurs when REF data measurement is aborted.
48
REF measurement incomplete
This error occurs when the cable correction REF measurement is incomplete.
S
82
Store failed
This error occurs when external mass storage device fails or internal FLASH ROM
hardware fails.
Contact Keysight Technology’s Sales and Service Office or the company from which you
purchased the device.
-102
Syntax error
There is a command or data type that cannot be recognized. For example, in the program
message “
:SYST:PRES
”, a colon (:) is inserted by mistake to give “
:SYST::PRES
”.
-134
Suffix too long
The length of suffix is long.
-138
Suffix not allowed
A suffix is added to a numeric value element that does not permit a suffix.
-150
String data error
An error not included in the error numbers between -151 and -159 has occurred during the
syntax analysis of a string data element.
-158
String data not allowed
A string data element has been received where the E4981A does not accept any string data
element. For example, a parameter must be enclosed with double quotation marks (“...”)
but they are missing.
T
-124
Too many digits
The number of digits of the mantissa of the decimal value data element exceeds 255 except
for preceding 0s. (Refer to IEEE488.27.7.2.4.1.)
Summary of Contents for E4981A
Page 1: ...Keysight E4981A 120 Hz 1 kHz 1 MHz Capacitance Meter Service Guide ...
Page 2: ......
Page 13: ...Contents 11 L 145 M 145 N 145 O 146 P 146 Q 146 R 147 S 147 T 147 U 148 Warning Message 149 ...
Page 14: ...12 Contents ...
Page 70: ...68 Chapter4 Troubleshooting Performance test failure troubleshooting ...
Page 84: ...82 Chapter5 Replaceable Parts Replaceable Parts List Front Panel Figure 5 12 Front Panel ...
Page 132: ...130 Chapter7 Post Repair Procedures Post Repair Procedures ...
Page 138: ...136 AppendixB Firmware Update Update the E4981A firmware using USB cable ...
Page 142: ...140 AppendixC Power Requirement Power Requirements ...