Chapter 4
51
Troubleshooting
Troubleshooting Using Service Function
4. T
roub
lesho
o
ting
The file location on the USB memory is as follows:
\e4981a\system\sysinfo.txt
NOTE
The system information of E4981A may be used for the purposes of support and repair by
Keysight Technologies, but will never be used for any other purposes.
To save the system information into external memory:
Step 1.
Plug a USB memory stick into the front USB port.
Step 2.
Press [System].
Step 3.
Press the
SERVICE
softkey.
Step 4.
Press the
MORE
softkey.
Step 5.
Press the
SAVE SYS INFO
softkey.
Step 6.
While the data is being saved, a "PROCESSING..." message is shown on the display. When
the data has been saved into the USB memory, a "Storing data completed." message
appears in the system message area.
Example of the System Information
The sysinfo.txt can be read by using an external computer (PC). The example of the system
information is as follows:
Example 4-1
System information
ID: Keysight Technologies,E4981A,MY12345678,A.01.00
OPT: 0
POWER CYCLE: 214
POWER ON HOURS: 746
SYSCAL REV: 1.00
SYSCAL DATE: 2008/7/23
R: Rev.1.11
CPU[1]: 0
CPU[2]: 0
SYST[1]: 0
SYST[2]: 0
SYST[3]: 0
SYST[4]: 0
DATA[1]: 0
DATA[2]: 0
DATA[3]: 0
DATA[4]: 0
OPT[1]: 0
A1[1]: -1.15601
A1[2]: +1.18401
A1[3]: +5.12900
A1[4]: +3.36400
Summary of Contents for E4981A
Page 1: ...Keysight E4981A 120 Hz 1 kHz 1 MHz Capacitance Meter Service Guide ...
Page 2: ......
Page 13: ...Contents 11 L 145 M 145 N 145 O 146 P 146 Q 146 R 147 S 147 T 147 U 148 Warning Message 149 ...
Page 14: ...12 Contents ...
Page 70: ...68 Chapter4 Troubleshooting Performance test failure troubleshooting ...
Page 84: ...82 Chapter5 Replaceable Parts Replaceable Parts List Front Panel Figure 5 12 Front Panel ...
Page 132: ...130 Chapter7 Post Repair Procedures Post Repair Procedures ...
Page 138: ...136 AppendixB Firmware Update Update the E4981A firmware using USB cable ...
Page 142: ...140 AppendixC Power Requirement Power Requirements ...