Chapter 7
129
Post-Repair Procedures
Post-Repair Procedures
7. Po
st-
R
e
p
air Pr
ocedu
re
s
LCD Assembly
No adjustment needed
Inspect the booting process, and power on
test result.
Key Assembly
No adjustment needed
Inspect the booting process, and power on
test result.
“To Check the Front Panel” on page 56
Interface Board
No adjustment needed
Inspect the booting process, and power on
test result.
“To Check the Handler Interface Function”
on page 60
“To Check the Scanner Interface Function”
on page 64
Table 7-1
Post-Repair Procedures
Replaced
Assembly or Part
Required Adjustments
Correction Constants (CC)
Verification