138
Appendix C
Power Requirement
Replacing Fuse
Replacing Fuse
Fuse Selection
Select proper fuse according to the
Setting up the fuse
Use the fuse that meets the following specifications.
UL/CSA type, Slo-Blo, 5x20mm miniature fuse, 3A 250V (part number: 2110-1017)
Spare fuses are available from Keysight Technologies sales office. To check or replace the
fuse, disconnect the power cable and pull out the fuse holder (refer to
) towards
you (refer to
).
Figure C-1
Fuse holder and power cable socket
Figure C-2
Removing the fuse
Table C-1
Fuse Selection
Fuse Rating/Type
Fuse Part Number
3A 250Vac
UL/CSA type
Time Delay
2110-1017
㪼㪋㪐㪏㪇㪸㫊㪼㪈㪇㪊㪐
Power cable receptacle
Fuse holder
USE ONLY WITH A 250V FUSE