48
Chapter 4
Troubleshooting
Troubleshooting Using Service Function
Figure 4-3
SELF TEST Page
Choosing a Test Item
Functional Description
E4981A can run the following self tests:
The tests of SYSTEM, USER DATA, and BATTERY are also performed by the power on
test.
㪼㪋㪐㪏㪈㪸㫊㪼㪈㪇㪇㪍
Table 4-1
Test item
Description
SYSTEM
Checks the entire system, A1/A2/A3 boards, system
calibration data, and so on.
USER DATA
Checks the GPIB/LAN configurations, instrument
configuration information, instrument calibration data, and
scanner calibration data.
BATTERY
Checks the internal batteries.
KEY
Checks the font panel keys. (Visual checks only; no result
shown on screen.)
DISPLAY
Checks the LCD. (Visual checks only; no result shown on
screen.)
HANDLER
INTERFACE
Checks the handler interface. (Visual checks only; no result
shown on screen.)