Chapter 2
27
Performance Test
Individual Test Description
3. Ch
apter
T
itle
4. Chapter
T
itle
5. Chapter T
itle
2. Pe
rfor
ma
nce T
e
st
Test point and limits
Cable Length : 2 m, MEAS TIME : 1
Range : 100 nF
Cable Length : 2 m, MEAS TIME : 8
Range : 100 nF
Cable Length : 2 m, MEAS TIME : 8
Range : 220 nF
Cable Length : 2 m, MEAS TIME : 8
Range : 470 nF
Standard
Signal
Level
Freq
Freq
Shift
Param
Test Limit
Measurement
Uncertainty
TAR
100 nF
1 V
1 kHz
0 %
Cp
±
85 pF
±
5.44 pF
15.63
0 %
D
±
0.00065
±
0.000032
20.83
Standard
Signal
Level
Freq
Freq
Shift
Param
Test Limit
Measurement
Uncertainty
TAR
100 nF
1 V
120 Hz
0 %
Cp
±
70 pF
±
7.68 pF
9.12
0 %
D
±
0.0005
±
0.000035
14.44
1 kHz
0 %
Cp
±
70 pF
±
4.37 pF
16.02
0 %
D
±
0.0005
±
0.000023
22.45
0.1 V
1 kHz
0 %
Cp
±
205 pF
±
12.36 pF
16.59
0 %
D
±
0.0019
±
0.000023
83.04
Standard
Signal
Level
Freq
Freq
Shift
Param
Test Limit
Measurement
Uncertainty
TAR
100 nF
1 V
1 kHz
0 %
Cp
±
88 pF
±
4.60 pF
19.13
0 %
D
±
0.00068
±
0.000023
30.54
Standard
Signal
Level
Freq
Freq
Shift
Param
Test Limit
Measurement
Uncertainty
TAR
100 nF
1 V
1 kHz
0 %
Cp
±
126 pF
±
7.10 pF
17.67
0 %
D
±
0.0011
±
0.000023
47.37
Summary of Contents for E4981A
Page 1: ...Keysight E4981A 120 Hz 1 kHz 1 MHz Capacitance Meter Service Guide ...
Page 2: ......
Page 13: ...Contents 11 L 145 M 145 N 145 O 146 P 146 Q 146 R 147 S 147 T 147 U 148 Warning Message 149 ...
Page 14: ...12 Contents ...
Page 70: ...68 Chapter4 Troubleshooting Performance test failure troubleshooting ...
Page 84: ...82 Chapter5 Replaceable Parts Replaceable Parts List Front Panel Figure 5 12 Front Panel ...
Page 132: ...130 Chapter7 Post Repair Procedures Post Repair Procedures ...
Page 138: ...136 AppendixB Firmware Update Update the E4981A firmware using USB cable ...
Page 142: ...140 AppendixC Power Requirement Power Requirements ...