30
Chapter 2
Performance Test
Individual Test Description
Cable Length : 2 m, MEAS TIME : 8
Range : 10 nF
Cable Length : 2 m, MEAS TIME : 8
Range : 10 pF
Cable Length : 2 m, MEAS TIME : 8
Range : 10 uF
Standard
Signal
Level
Freq
Freq
Shift
Param
Test Limit
Measurement
Uncertainty
TAR
10 nF
1 V
120 Hz
0 %
Cp
±
7 pF
±
0.61 pF
11.60
0 %
D
±
0.0005
±
0.000024
20.89
1 kHz
0 %
Cp
±
7 pF
±
0.71 pF
9.94
0 %
D
±
0.0005
±
0.000021
24.83
Standard
Signal
Level
Freq
Freq
Shift
Param
Test Limit
Measurement
Uncertainty
TAR
10 pF
1 V
1 MHz
*1
-2 %
Cp
±
7 fF
±
1.11 fF
6.34
-2 %
D
±
0.0005
±
0.000065
7.73
-1 %
Cp
±
7 fF
±
1.14 fF
6.15
-1 %
D
±
0.0005
±
0.000065
7.73
0 %
Cp
±
7 fF
±
0.98 fF
7.16
0 %
D
±
0.0005
±
0.000065
7.73
1 %
Cp
±
7 fF
±
0.75 fF
9.34
1 %
D
±
0.0005
±
0.000065
7.73
2 %
Cp
±
7 fF
±
0.84 fF
8.40
2 %
D
±
0.0005
±
0.000065
7.73
*1.Option 001 only.
Standard
Signal
Level
Freq
Freq
Shift
Param
Test Limit
Measurement
Uncertainty
TAR
10 uF
1 V
120 Hz
0 %
Cp
±
7 nF
±
0.53 nF
13.40
0 %
D
±
0.0005
±
0.000072
6.98
1 kHz
0 %
Cp
±
7 nF
±
0.79 nF
8.96
0 %
D
±
0.0005
±
0.000066
7.67
Summary of Contents for E4981A
Page 1: ...Keysight E4981A 120 Hz 1 kHz 1 MHz Capacitance Meter Service Guide ...
Page 2: ......
Page 13: ...Contents 11 L 145 M 145 N 145 O 146 P 146 Q 146 R 147 S 147 T 147 U 148 Warning Message 149 ...
Page 14: ...12 Contents ...
Page 70: ...68 Chapter4 Troubleshooting Performance test failure troubleshooting ...
Page 84: ...82 Chapter5 Replaceable Parts Replaceable Parts List Front Panel Figure 5 12 Front Panel ...
Page 132: ...130 Chapter7 Post Repair Procedures Post Repair Procedures ...
Page 138: ...136 AppendixB Firmware Update Update the E4981A firmware using USB cable ...
Page 142: ...140 AppendixC Power Requirement Power Requirements ...