34
Chapter 2
Performance Test
Individual Test Description
Cable Length : 0 m, MEAS TIME : 8
Range : 100 pF
Cable Length : 2 m, MEAS TIME : 8
Range : 1 mF
Cable Length : 2 m, MEAS TIME : 8
Range : 0.1 mF
Cable Length : 2 m, MEAS TIME : 8
Range : 0.1 mF
Standard
Signal
Level
Freq
Freq
Shift
Param
Test Limit
Measurement
Uncertainty
TAR
100 pF
1 V
1 MHz
*1
-2 %
Cp
±
70 fF
±
10.04 fF
6.97
-2 %
D
±
0.0005
±
0.000065
7.71
-1 %
Cp
±
70 fF
±
10.20 fF
6.86
-1 %
D
±
0.0005
±
0.000065
7.71
0 %
Cp
±
70 fF
±
10.83 fF
6.46
0 %
D
±
0.0005
±
0.000065
7.71
1 %
Cp
±
70 fF
±
10.50 fF
6.66
1 %
D
±
0.0005
±
0.000065
7.71
2 %
Cp
±
70 fF
±
10.31 fF
6.79
2 %
D
±
0.0005
±
0.000065
7.71
*1.Option 001 only.
Standard
Signal
Level
Freq
Freq
Shift
Param
Test Limit
Measurement
Uncertainty
TAR
1
Ω
1 V
120 Hz
0 %
Rs
±
4.3 m
Ω
±
0.15 m
Ω
29.48
Standard
Signal
Level
Freq
Freq
Shift
Param
Test Limit
Measurement
Uncertainty
TAR
1
Ω
1 V
1 kHz
0 %
Rs
±
4.3 m
Ω
±
0.15 m
Ω
30.40
Standard
Signal
Level
Freq
Freq
Shift
Param
Test Limit
Measurement
Uncertainty
TAR
10
Ω
1 V
120 Hz
0 %
Rs
±
7.0 m
Ω
±
1.30 m
Ω
5.39
Summary of Contents for E4981A
Page 1: ...Keysight E4981A 120 Hz 1 kHz 1 MHz Capacitance Meter Service Guide ...
Page 2: ......
Page 13: ...Contents 11 L 145 M 145 N 145 O 146 P 146 Q 146 R 147 S 147 T 147 U 148 Warning Message 149 ...
Page 14: ...12 Contents ...
Page 70: ...68 Chapter4 Troubleshooting Performance test failure troubleshooting ...
Page 84: ...82 Chapter5 Replaceable Parts Replaceable Parts List Front Panel Figure 5 12 Front Panel ...
Page 132: ...130 Chapter7 Post Repair Procedures Post Repair Procedures ...
Page 138: ...136 AppendixB Firmware Update Update the E4981A firmware using USB cable ...
Page 142: ...140 AppendixC Power Requirement Power Requirements ...