Front Panel Operation
2-69
when the Model 6517A receives a trigger via the selected
trigger link line. Note that the TRIG key can instead be used
to start the test.
Table 2-20
CONFIGURE SEQUENCE menu structure
Menu item
Description
Default
parameter
APPLICATIONS
DEV-CHAR
DIODE
LEAKAGE-CURRENT
START V
STOP V
STEP V
DELAY
CAPACITOR
LEAKAGE-CURRENT
BIAS V
STORE nnnnn READINGS
INTERVAL
CABLE
INSULATION-RESISTANCE
BIAS V
STORE READINGS
INTERVAL
RESISTOR
VOLTAGE-COEFFICIENT
SOURCE V1
DELAY 1
SOURCE V2
DELAY 2
Select type of test:
Device Characterization Tests:
Diode Leakage Current Test:
Specify start voltage.
Specify stop voltage.
Specify step voltage.
Specify delay.
Capacitor Leakage Current Test:
Specify bias voltage.
Specify number of readings.
Specify time interval.
Cable Insulation Resistance Test:
Specify bias voltage.
Specify number of readings.
Specify time interval.
Resistor Voltage Coefficient Test:
Specify 1st test voltage.
Specify 1st delay.
Specify 2nd test voltage.
Specify 2nd delay.
+1V
+10V
+1V
1sec
+1V
10
1 sec
+1V
5
1 sec
+1V
1sec
+2V
1 sec
R/RESISTIVITY
NORMAL
SURFACE
PRE-DISCH
BIAS V
BIAS-TIME
MEAS-V
MEAS-TIME
DISCHARGE
VOLUME
PRE-DISCH
BIAS V
BIAS-TIME
MEAS-V
MEAS-TIME
DISCHARGE
Resistance/Resistivity Tests:
Standard Method Resistivity Tests:
Surface Resistivity Test:
Specify pre-discharge time.
Specify bias voltage.
Specify bias time.
Specify measurement voltage.
Specify measurement time.
Specify discharge time.
Volume Resistivity Test:
Specify pre-discharge time.
Specify bias voltage.
Specify bias time.
Specify measurement voltage.
Specify measurement time.
Specify discharge time.
0.2 sec
+500V
1 sec
+500V
0 sec
2 sec
10 sec
+500V
1 sec
+500V
0 sec
2 sec
TRIGLINK: Use this option to select the trigger link trigger
source. After selecting TRIGLINK you will be prompted to
select the trigger link line. Once the test is armed, it will start