![background image](http://html2.mh-extra.com/html/keithley/6517a/6517a_user-manual_4066185214.webp)
IEEE-488 Reference
3-63
:MA2Factor <NRf>
:CALCulate[1]:KMATh:MA2Factor <NRf> Specify "a2" factor
Parameters
<NRf> = -9.999999e20 to +9.999999e20
Query
:MA2Factor?
Query "a2" factor
Description
This command is used to specify the "a2" factor for the polynomial calculation.
:PERCent <NRf>
:CALCulate[1]:KMATh:PERCent <NRf> Specify Percent target value
Parameters
<NRf> = -9.999999e35 to +9.999999e35
Specify target value for Percent calculation.
Query
:PERCent?
Query Percent target value
Description
This command is used to specify the target value for the Percent calculation.
:REFerence <NRf>
:CALCulate[1]:REFerence <NRf> Set reference value for ratio, deviation and percent deviation
Parameters
<NRf> = -9.999999e35 to +9.999999e35
Specify reference value
Query
:REFerence?
Query reference value
Description
This command is used to specify the reference value for the ratio, deviation and percent devia-
tion calculation.
:STATe <b>
:CALCulate[1]:STATe <b> Control CALC 1
Parameters
<b> = 0 or OFF
Disable CALC 1 calculation
1 or ONEnable CALC 1 calculation
Query
:STATe?
Query state (on or off) of CALC 1
Description
This command is used to enable or disable the CALC 1 calculation. When enabled, each instru-
ment reading will reflect the selected calculation (see :FORMat).
:DATA commands
[:LATest]?
:CALCulate[1]:DATA[:LATest]? Read CALC 1 result
Description
This query command is used to read the result of the CALC 1 calculation. If CALC 1 is disabled
or NONE is selected, the "raw" reading will instead be read.