Front Panel Operation
2-17
To 6517A
Interlock
To 6517A
V Source
Out
To 6517A
Input
Interlock Switch
Open Lid = Open Switch
Insulated
Terminal
Post (5)
Screw Terminal for
Safety Earth Ground
Warning
: Test fixture must
be connected to safety earth
ground using #18 AWG wire
or larger.
Interlock Connector
Dual Banana Jacks
3-Lug Female Triax Connector
Banana Jack
A
B
C
Guard Plate
1
2
3
A
B
C
DUT
or
Test
Circuit
D
To 6517A
Common
D
4
Figure 2-17
Multi-purpose test fixture
6517A
A. Interlock Connection to Unit
Side View
View from inside
of test box
Normally-open SPST
Momentary Switch
Open lid = Open Switch
D. Interlock Wiring
C. Interlock Connector:
Keithley P/N : CS-659 (3-pin)
CS-459 (4-pin)
Switchcraft P/N :
TB3M (3-pin)
TB4M (4-pin)
.444
+.004
-.000
.196
.422
+.004
-.000
B. Panel cutout dimensions
#3
#1
#2
#4
Interlock
Interlock
Connector
Interlock Cable
Test Fixture
Figure 2-18
Interlock connections