IEEE-488 Reference
3-102
3.19.14 :ADIScharge Commands
Auto discharge is used to reset the charge reading to zero. See paragraph 2.8 for more informa-
tion.
[:STATe] <b>
[:SENSe[1]]:CHARge:ADIScharge[:STATe] <b>
Control auto discharge
Parameters
<b> =
1 or ON
Enable auto discharge
0 or OFF
Disable auto discharge
Query
[:STATe]?
Query state of auto discharge
Description
This command is used to enable or disable the auto discharge feature for the Coulombs function.
When enabled, the charge measurement will reset to zero when the specified level is reached
(see next command).
:LEVel <NRf>
[:SENSe[1]]:CHARge:ADIScharge:LEVel <NRf>
Specify level
Parameters
<NRf> = -2.2e6 to 2.2e6
Specify discharge level.
Query
:LEVel?
Query level for auto discharge
Description
This command is used to specify the coulombs level for auto discharge. Charge will reset when
the charge reading reaches the specified level.
3.19.15 :XFEedback <b>
[:SENSe[1]]:VOLTage[:DC]:XFEedback <b>
Parameters
<b> = ON or 1
Enable external feedback
OFF or 0
Disable external feedback
Query
:XFEedback?
Query state of external feedback
Description
This command is used to enable or disable external feedback. See paragraph 2.11 for informa-
tion on external feedback.
3.19.16 :VSControl <name>
[:SENSe[1]]:RESistance:VSControl <name>
Select ohms V-Source mode
Parameters
<name> = MANual
Manual V-Source Ohms
AUTO
V-Source Ohms
Query
:VSControl?
Query ohms V-Source mode.
Description
This command is used to select Manual V-Source ohms or Auto V-Source ohms. With MANual
selected, you can set the V-Source level for ohms measurements. See paragraph 3.19.6 (Ohms
Ranges) to configure and control the V-Source for Manual V-Source ohms measurements.
With AUTO selected, the V-Source level is automatically set for the ohms measurement (40V
or 400V).