:SSYNchronous:LINE <NRf>
:TRIGger[:SEQuence[1]]:TCONfigure:SSYNchronous :LINE <NRf>
Specify semi-synchronous Trigger
Link line for measure layer
Parameters
<NRf> = 1 Line #1
<NRf> = 4 Line #4
2 Line #2
5 Line #5
3 Line #3
6 Line #6
Query
:LINE?
Query programmed output line #
Parameters
This command is used to select one of the six trigger lines for the semi-synchronous Trigger
Link.
3.24.10 RTCLock commands
The following commands are used to specify the date and time for the RTCLock control source
(see paragraph 3.24.6 for details).
:DATE <yr>, <mo>, <day>
:ARM[:SEQuence[1]][LAYer[]]:RTCLock:DATE <yr>, <mo>, <day>
Set date for RTCLock control source
Parameters
<yr> = 1994 to 2093
Specify year
<mo> = 1 to 12
Specify month
<day> = 1 to 31
Specify day
Query
:DATE?
Query the date
Description
This command is used to set the date for the RTCLock control source. Setting an invalid date
(i.e. February 29, 1994) for the clock event results in an error and the previous date is retained.
:TIME <hr>, <min>, <sec>
:ARM[:SEQuence[1]][LAYer[]]:RTCLock:TIME <hr>, <min>, <sec>
Set time for RTCLock control source
Parameters
<hr> = 0 to 23
Specify hour (24 hour format)
<min> = 0 to 59
Specify minute
<sec> = 0.00 to 59.99
Specify second (round to 1/100 sec.)
Query
:DATE?
Query the time
Description
This command is used to set the time for the RTCLock control source. Note that the time must
be set using the 24-hour format (i.e. hour 13 is 1pm). Setting an invalid time results in an error
and the previous time is retained.
:TIME? returns the time to nearest hundredth of a second.