Front Panel Operation
2-72
Table 2-21 (cont.)
CONFIGURE TRIGGER menu structure
Menu item
Description
SCAN
SOURCE
IMMEDIATE
EXTERNAL
MANUAL
GPIB
TRIGLINK
TIMER
HOLD
DELAY
COUNT
INFINITE
ENTER-SCAN-COUNT
CONTROL
SOURCE
ACCEPTOR
Scan layer menu:
Select scan source:
Use to pass operation immediately into the measure layer.
Use external triggers to control scanning.
Use TRIG key to control scanning.
Use bus triggers to control scanning.
Use Trigger Link triggers to control scanning. Enter Trigger Link lines.
Use a timer to control scanning and enter interval between scans (0.001 -
999999.999 sec.).
Use to hold up the measurement in the scan layer.
Use to delay scan in the layer (0.001 - 999999.999 sec.).
Define number of scans to be performed:
Repeat scanning indefinitely.
Specify count (1 - 99999).
Select trigger control mode:
Enable Source Bypass.
Disable Source Bypass.
ARM
SOURCE
IMMEDIATE
EXTERNAL
MANUAL
GPIB
TRIGLINK
RT-CLOCK
HOLD
COUNT
INFINITE
ENTER-ARM-COUNT
CONTROL
SOURCE
ACCEPTOR
Arm layer menu:
Select arm source:
Use to arm meter immediately and pass operation into the scan layer.
Use external triggers to arm meter.
Use TRIG key to arm meter.
Use bus triggers to arm meter.
Use Trigger Link triggers to arm meter. Enter Trigger Link lines.
Use clock to arm instrument. Enter time and date.
Use to hold up the measurement in the arm layer.
Define number of times to arm meter:
Continuously re-arm meter.
Specify count (1 - 99999).
Select trigger control mode:
Enable Source Bypass.
Disable Source Bypass.
INIT
HALT
Enable or disable continuous initiation.
Use to halt triggers. Press TRIG key to resume triggering.