Device Dependent Command Summary
G-2
Reading Mode
B0
B1
B2
B3
B4
Electrometer
Buffer reading
Maximum reading
Minimum reading
Voltage source
Data Store
Q0
Q1
Q2
Q3
Q4
Q5
Q6
Q7
Conversion rate
One reading per second
One reading every 10 seconds
One reading per minute
One reading every 10 minutes
One reading per hour
Trigger mode
Disabled
Voltage Source Value
V±nnnn.nn or
V±n.nnnnE±n
Voltage source value: -1000V to +1000V, 50mV increments
Voltage Source Operate
O0
O1
Source output off (0V)
Source output on (programmed value)
Data Format
G0
G1
G2
Reading with prefix (NDCV -1.00)
Reading without prefix (-1.00)
Reading with prefix and buffer suffix (if in B1)
(NDCV -1.00, 012)
Trigger Mode
T0
T1
T2
T3
T4
T5
T6
T7
Continuous, trigger by talk
One-shot, trigger by talk
Continuous, trigger by GET
One-shot, trigger by GET
Continuous, trigger by X
One-shot, trigger by X
Continuous, trigger by external trigger
One-shot, trigger by external trigger
SRQ
M0
M1
N2
M8
M16
M32
Disable SRQ
Reading overflow
Buffer full
Reading done
Ready
Error
EOI and Bus Hold Off
K0
K1
K2
K3
Enable EOI and bus hold-off on X
Disable EOI, enable bus hold-off on X
Enable EOI, disable bus hold-off on X
Disable both EOI and bus hold-off on X
Terminator
Y0
Y1
Y2
Y3
Y4
LFCR
CRLF
LF
CR
None
Mode
Command
Description