IEEE-488 Reference
3-46
Table 3-9
ROUTe command summary
Command
Description
Default
parameter
SCPI
Ref.
:ROUTe
:CLOSe <list>
:STATe?
:CLOSe? <list>
:OPEN <list>
:OPEN:ALL
:OPEN? <list>
:SCAN
[:INTernal] <list>
[:INTernal]?
:EXTernal <n>
:EXTernal?
:LSELect <name>
:LSELect?
:STIMe <n>
:STIMe?
:SMEThod <name>
:SMEThod?
:VSLimit <b>
:VSLimit?
Path and command to close specified channel:
Query closed channel.
Query specified channels (1 = closed, 0 = open).
Open specified channel or all channels.
Open all channels.
Query specified channels (1 = open, 0 = closed).
Command path to configure and control scan operations:
Specify internal scan list (2 to 10 channels).
Query the internal scan list.
Specify external scan list (1 to 400 channels).
Query the external scan list.
Select scan operation: (INTernal, EXTernal, or
NONE).
Query scan operation.
Specify settling time for internal card (0 to 99999.9999
seconds)
Query settling time.
Select scan method for internal scan (VOLTage or
CURRent).
Query scan method.
Enable or disable 200V limit for internal card.
Query state of 200V limit
All 10
10
NONE
0
VOLTage
ON
√
√
√
√
√
√
√
√
√
3.18.1
3.18.2
3.18.3
3.18.4
Table 3-10
SENSe command summary
Command
Description
Default
parameter
SCPI
Ref.
[:SENSe[1]]
:FUNCtion <name>
:FUNCtion?
:DATA
[:LATest]?
:FRESh?
Select measurement function: ‘VOLTage[:DC]’,
‘CURRent[:DC]’, ‘RESistance’ , ‘CHARge’
Query function.
Path to returnreadings:
Return the last instrument reading.
Query fresh readings.
‘VOLT:DC’
√
√
√
√
√
3.19.1
3.19.2
3.19.3