IEEE-488 Reference
3-131
The :KEY command allows you to select the NEXT displays over the bus. Sending :SYS-
Tem:KEY 16 over the bus is same as a single press of the NEXT key. Sending :SYSTem:KEY
8 is the same as a single press of the PREV key. The NEXT display that is selected depends on
how many times the NEXT key is "pressed" using the :KEY command. The instrument can al-
ways be returned to the normal display by "pressing" the PREV key a similar number of times,
or by sending the :DISPlay:CNDisplay command.
For example, you can display the bar graph by "pressing" the NEXT key twice as follows:
:syst:key 16; key 16
To return to the normal display, "press" the PREV key twice as follows:
:syst:key 8; key 8
The queue for the :KEY? query command can only hold one key-press. When :KEY? is sent
over the bus, and the Model 6517A is addressed to talk, the key-press code number for the last
key pressed (either physically or with :KEY) is sent to the computer.
3.22.7
:CLEar
:SYSTem:CLEar
Clear Error Queue
Description
This action command is used to clear the Error Queue of messages.
EXIT
ENTER
CONFIG MENU
TRIG
SEQ
STORE RECALL
REL
Z-CHK
INFO
LOCAL
V
I
R
Q
CARD
OPER
FILTER MATH
OPTION
VOLTAGE SOURCE
6517 ELECTROMETER/HIGH RESISTANCE METER
RANGE
VOLTAGE
SOURCE
OPERATE
AUTO
PREV
NEXT
DISPLAY
POWER
RANGE
CAL
1
2
3
4
5
6
7
8
9
10
1
2
3
4
5
6
7
8
9
10
EDIT ERR REM TALK
LSTN SRQ REAR
REL
FILT MATH
4W AUTO
ARM TRIG
SMPL
8
15
22
29
5
12
19
26
2
1
16
23
31
7
30
6
14
21
13
20
28
4
27
3
11
18
10
17
9
6517A ELECTROMETER/HIGH RESISTANCE METER
Figure 3-40
Key-press codes