Basic Service Manual
8 - 234
Section 8-12 - Functional Checks after FRU replacement (Debrief Guideline)
8-12-69
Replacement of the UCO-T Assy
8-12-70
Replacement of the Unity ACDC Assy
8-12-71
Replacement of the UHUB-T
Functional Check
Debrief Script
Leakage Current measured at (record the value) and
meets allowable limits. Equipment passed all required
checks and is ready for use.
4-3-7 "Probe/connector functional checks"
4-3-8 "B-Mode functional checks"
4-3-10 "System CFM and PWD functional
4-3-15 "CW Doppler mode functional checks"
4-3-1 "Control panel adjustment and functional
10-6-6 "Earth Leakage Current Test"
Functional Check
Debrief Script
Leakage Current measured at (record the value) and
meets allowable limits. Equipment passed all required
checks and is ready for use.
4-3-7 "Probe/connector functional checks"
4-3-8 "B-Mode functional checks"
4-3-25 "Battery Pack functional checks"
4-3-26 "Battery Pack Extended functional checks"
(if the system has Battery Pack)
10-6-6 "Earth Leakage Current Test"
Functional Check
Debrief
Script
Equipment passed all required checks and is ready for
use.
4-3-7 "Probe/connector functional checks"
4-3-8 "B-Mode functional checks"
4-3-10 "System CFM and PWD functional checks"
Summary of Contents for LOGIQ Totus
Page 2: ......
Page 4: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual ii...
Page 5: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual iii...
Page 6: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual iv...
Page 643: ......