GE HealthCare CONFIDENTIAL
D
IRECTION
5936427
, R
EV
. 1
LOGIQ T
OTUS
Basic Service Manual
Section 8-4 - Loading / Reloading / Upgrading the Software
8 - 29
8-4-9
Loading only the application software
Follow this procedure to only load the application software (and not the Windows operating system).
If disk encryption is enabled, you will need the correct recovery keys (32-bit string) and password (if set
up). If recovery keys are not available, you will have to do a full load instead. The only way to recover
the patient data is if the customer has previously backed up data and images.
NOTE:
Lockout/tagout (LOTO) is not required as you will not manipulate any electronics.
NOTE:
A Service key is not required to run this procedure.
This table includes information specific to these instructions. For more information
for servicing the LOGIQ Totus" on page 8-5
8-2-7 "PPE Required During Service" on page 8-6
Table 8-11 Manpower, Time, Tools, and PPE
Manpower
Time
Tools
PPE
One person
Approximately 1 hour and 30 minutes
to 2 hours and 30 minutes depending
on the amount of back-up/restore
activities (presets and patient archive)
needed prior to and after the
installation.
Phillips screw driver
Software thumb drive
No PPE required