GE HealthCare CONFIDENTIAL
D
IRECTION
5936427
, R
EV
. 1
LOGIQ T
OTUS
Basic Service Manual
Section 8-8 - Replacing Main Console
8 - 139
8-8-11 Replacement of the Air Filter Set-U
This table includes information specific to these instructions. For more information
for servicing the LOGIQ Totus" on page 8-5
8-2-7 "PPE Required During Service" on page 8-6
8-8-11-1
Preparations
1.) Remove the T2 Side Cover L Assy as described in
8-8-7 "Replacement of the T2 Side Cover L
2.) Remove the T2 Side Cover R Assy as described in
8-8-8 "Replacement of the T2 Side Cover R
3.) Remove the Rear Cover Mid Assy-T as described in
8-8-10 "Replacement of the Rear Cover Mid
8-8-11-2
Removal Procedure
8-8-11-3
Installation Procedure
1.) Parts to be installed in reverse order of removal
2.) Perform “Functional Checks after FRU replacement (Debrief Guideline)” as described in
Table 8-73 Manpower, Time, Tools, and PPE
Manpower
Time
Tools
PPE
One person
Approximately 5 minutes
No tools required
Cut resistant gloves
Knee Pad
Step
Corresponding graphic
1. Release 2 snap fits to remove the Unity Air Filter Side
from the T2 Side Cover L/R Assy.
2. Release 3 snap fits to remove the Unity Air Filter Rear
from the Rear Cover Mid Assy-T.
Summary of Contents for LOGIQ Totus
Page 2: ......
Page 4: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual ii...
Page 5: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual iii...
Page 6: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual iv...
Page 643: ......