, Rev. 1
LOGIQ Totus
Basic Service Manual
Chapter 10 Care & Maintenance
10 - 7
Console Leakage
Current Checks
•
Also after corrective
maintenance or as required by
your facilities QA program.
Peripheral Leakage
Current Checks
•
Also after corrective
maintenance or as required by
your facilities QA program.
Surface Probe Leakage
Current Checks
•
Also after corrective
maintenance or as required by
your facilities QA program.
Endocavity Probe
Leakage
Current Checks
•
Also after corrective
maintenance or as required by
your facilities QA program.
Measurement Accuracy
Checks
•
Also after corrective
maintenance or as required by
your facilities QA program.
Probe/Phantom Checks
Check axial and lateral resolution ( Basic
User Manual Technical specifications).
Check Gain and TGC changes, vary the
focus and check reaction on screen. Check
deviation of brightness in the US-Image
(missing elements / probe cable defect).
Probe must be coupled as this test.
•
Also after corrective
maintenance or as required by
your facilities QA program.
Functional Checks of All
Probes
Check general functions and image
appearance at human body with all
available modes.
•
Also after corrective
maintenance or as required by
your facilities QA program.
Table 10-1 Customer Care Schedule (Continued)
Item
Service at Indicated Time
D
W
M
A
Notes
Summary of Contents for LOGIQ Totus
Page 2: ......
Page 4: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual ii...
Page 5: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual iii...
Page 6: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual iv...
Page 643: ......