GE HealthCare CONFIDENTIAL
D
IRECTION
5936427
, R
EV
. 1
LOGIQ T
OTUS
Basic Service Manual
Section 8-8 - Replacing Main Console
8 - 157
8-8-25 Replacement of the Unity Top Fan Assy
This table includes information specific to these instructions. For more information
for servicing the LOGIQ Totus" on page 8-5
8-2-7 "PPE Required During Service" on page 8-6
8-8-25-1
Preparations
1.) Remove the T2 Side Cover R Assy as described in
8-8-8 "Replacement of the T2 Side Cover R
2.) Remove the Top cover Assy-T as described in
8-8-17 "Replacement of the Top Cover Assy-T" on
8-8-25-2
Removal Procedure
Table 8-87 Manpower, Time, Tools, and PPE
Manpower
Time
Tools
PPE
One person
Approximately 10 minutes
Phillips Screwdriver
Cut resistant gloves
Knee Pad
Step
Corresponding graphic
1.
Unclamp the cables and disconnect the Fan
Cable from the console.
2.
Unscrew 4 screws to remove the Unity Top
Fan Assy from the console.
Summary of Contents for LOGIQ Totus
Page 2: ......
Page 4: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual ii...
Page 5: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual iii...
Page 6: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual iv...
Page 643: ......