Section 8-11 - Replacing Options
8 - 207
8-11-17-2
Removal Procedure
8-11-17-3
Installation Procedure
1.) Parts to be installed in reverse order of removal
2.) Perform “Functional Checks after FRU replacement (Debrief Guideline)” as described in
Steps
Corresponding Graphic
1.
Push the button of Unity Protector Assy’s connector
and disconnect the Unity Protector Assy’s Connector
from the console.
2.
Unscrew 6 screws to remove the Unity UCO-T Front
Cover Brkt..
3.
Unscrew 2 screws to remove the U-DCWD Set from
the console.