GE HealthCare CONFIDENTIAL
Direction
5936427
, Rev. 1
LOGIQ Totus
Basic Service Manual
4 - 30
Section 4-3 - Functional Checks
3.) Then the system should be able to be unplugged and remain on for at least several seconds and
plugged back on without shutting down. A message may appear to user with a notification of being
in this mode.
4.) Verify that Scan is continued while the system is unplugged without shutdown.
5.) Reconnect AC power. Verify that the batteries begin charging via the battery status icon.
4-3-27
Shear Wave Elastography (Shear Elasto) functional checks
1.) Select a probe (C1-6-D).
2.) Select the
ELASTO
button.
3.) From the touch panel, select the
Shear Wave
button (if not already selected).
4.) Verify that the
Shear Wave
button is available.