
GE HealthCare CONFIDENTIAL
Direction
5936427
, Rev. 1
LOGIQ Totus
Basic Service Manual
4 - 2
Section 4-1 - Overview
4-1-4
Recommended test phantoms
GE HealthCare recommends the RMI 430GS phantom is preferred, but it is not required. It is the most
current phantom recommended to our field service personnel and provides the necessary targets and
extended life necessary for consistent Ultrasound system testing.
Figure 4-1 Performance Tests
2
4
3
2
1
4
6
8
10
12
14
16
RMI 403GS
Lateral Distance
Measurement
Functional
Resolution
Contrast
Resolution
&
Gray Scale
Photography
Lateral
Resolution
Axial
Resolution
Axial Distance
Measurement
Penetration
Summary of Contents for LOGIQ Totus
Page 2: ......
Page 4: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual ii...
Page 5: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual iii...
Page 6: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual iv...
Page 643: ......