Basic Service Manual
8 - 68
Section 8-5 - Replacing Cables
8-5-12-2
Removal Procedure
Steps
Corresponding Graphic
1.
Unscrew 4 screws to remove the Unity Cable Cover
Brkt.
2.
Unscrew 4 screws to remove the Unity Cable Hide
Brkt from the console.
3.
Unlock the Cable clamp and disconnect the ACDC
Link Cable To UBE Cable from the Unity ACDC Assy.
4.
Disconnect the ACDC Link Cable To UBE Cable from
the UBE.