
GE HealthCare CONFIDENTIAL
Direction
5936427
, Rev. 1
LOGIQ Totus
Basic Service Manual
Chapter 7 Diagnostics/Troubleshooting
7 - 55
Section 7-7
Options Troubleshooting
Contents in this section
Unable to measure in both stiffness and velocity in Breast and Thyroid . . . . . . . . . . . 7-55
7-7-1
ECG display disappears after backup/restore
With ECG, after system auto-restart, turn system off and then back on. Otherwise ECG and TGC may
not function until next full power cycle.
7-7-2
Unable to measure in both stiffness and velocity in Breast and Thyroid
WIth Shearwave, when entering Measurement in Breast or Thyroid, only the stiffness measurements
show up on the touch panel if display units is set to kPa, even though both stiffness and velocity are
enabled in the Utility pages.
Summary of Contents for LOGIQ Totus
Page 2: ......
Page 4: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual ii...
Page 5: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual iii...
Page 6: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual iv...
Page 643: ......