Basic Service Manual
Section 8-8 - Replacing Main Console
8 - 153
8-8-23 Replacement of the AC Inlet Fuse 2EA
This table includes information specific to these instructions. For more information
for servicing the LOGIQ Totus" on page 8-5
8-2-7 "PPE Required During Service" on page 8-6
.
8-8-23-1
Preparations
1.) Power off/Shutdown the system as described in
1-7-2 "Lockout/tagout procedure" on page 1-25
2.) Remove the Unity Power Cord Fix Brkt as described in
8-8-21 "Replacement of the Unity Power
8-8-23-2
Removal Procedure
8-8-23-3
Installation Procedure
1.) Parts to be installed in reverse order of removal
2.) Perform “Functional Checks after FRU replacement (Debrief Guideline)” as described in
Table 8-85 Manpower, Time, Tools, and PPE
Manpower
Time
Tools
PPE
One person
Approximately 5 minutes
Phillips screwdriver
Cut resistant gloves
Knee Pad
Steps
Corresponding Graphic
1.
Pull the Power Cord to disconnect the Power Cord
from the console.
2.
Push side of the Drawer and pull the Drawer
3.
Pull 2 Fuses from the Drawer.