GE HealthCare CONFIDENTIAL
Direction
5936427
, Rev. 1
LOGIQ Totus
Basic Service Manual
10 - 12
Section 10-4 - System Maintenance
10-4-10
Physical inspection
NOTE:
These features may not be present on all Ultrasound systems.
Table 10-7 Physical Checks
Step
Item
Description
1
Labeling
Verify that all Ultrasound system labeling is present and in readable condition.
2
Scratches &
Dents
Inspect the exterior for dents, scratches or cracks.
3
Covers
Where applicable, verify that all covers are secured in place and are properly aligned with other
covers. Replace any covers that are damaged.
4
Input Power
10-4-7 "Mains cable inspection" on page 10-11
5
External I/O
Check all connectors for damage.
6
Wheels and
Brakes
• Where applicable, check all wheels and casters for wear and verify the operation of foot brake,
which stops the Ultrasound system from moving, and the release mechanism.
• Where applicable, check all wheel locks and wheel swivel locks for proper operation.
7
Control Panel
Movement
• Where applicable, verify the ease of Operator Panel (Operator Control Panel) movement in all
acceptable directions.
• Where applicable, ensure that the Control panel latches in position as required.
8
Control Panel
Lighting
Check for proper operation of all operator panel lights.
9
Display Monitor
Inspect the displays for scratches and bad pixels.
Verify the proper operation of Contrast and Brightness controls.
Where applicable, confirm that the monitor arm allows
• Swiveling the screen to the left and to the right.
• Folding the screen to the locked position.
• Can be adjusted in the up/down positions.
10
Cables and
Connectors
Check all internal cable harnesses and connectors for wear and secure connector seating. Pay
special attention to probe strain or bend reliefs.
11
Shielding and
Covers
Check to ensure that all EMI shielding, internal covers, air flow panels and screws are in place.
Missing covers and hardware could cause EMI/RFI problems while scanning.
12
Control Panel &
Keyboard
Inspect alphanumeric keyboard and operator panel. Record any damaged or missing items.
(Replace faulty components as necessary).
13
Probe Holders
Where applicable, inspect the Probe Holders for cracks or damage.
14
Power and
System Status
Indicators
Check for proper operation of all Power and System Status Indicators.
15
Batteries
Where applicable, check that the batteries are not damaged, do not leak, do not emit an odor, and
are not deformed or discolored. Observe all warnings and cautions for battery handling, recharging,
storing, and/or disposal.
Summary of Contents for LOGIQ Totus