Section 8-5 - Replacing Cables
8-5-10-2
Removal Procedure
Steps
Corresponding Graphic
1.
Unlock the Cable Clamp and disconnect the UBE To
OPIO HUB USB 3.0 Cable from the OPIO HUB.
2.
Unscrew 2 screws to remove the Cable Clamp.
3.
Cut 2 Cable Ties. Pull UBE To OPIO Touch Panel
HDMI Cable to remove from the T2 OPIO BASE.
4.
Unscrew 4 screws to remove the Unity Cable Cover
Brkt