GE HealthCare CONFIDENTIAL
D
IRECTION
5936427
, R
EV
. 1
LOGIQ T
OTUS
Basic Service Manual
Section 8-4 - Loading / Reloading / Upgrading the Software
8 - 33
Loading only the application software (continued)
Verifying the software versions:
1.) From the touch panel, select
Utility
>
System
>
About
.
2.) Verify software version matches the version in the software used during installation, version
received with the kit.
Verifying the TCP/IP settings:
1.) From the touch panel, select
Utility
>
Connectivity
>
TCP/IP
.
2.) Verify TCP/IP settings.
3.) If you modified or updated any settings, select
Save settings
.
4.) Select
OK
to continue.
5.) Reboot the system.
Verifying the AE Title and Port settings:
1.) From the touch panel, select
Utility
>
Connectivity
>
Device
.
2.) Verify device settings.
3.) If you modified or updated any settings, select
Save settings
.
4.) Select
OK
to continue.
Check patient archive
Confirm that the patient archive is available. If the patient archive is encrypted, you will need the
encryption key or password to unlock.
Check system configurations
1.) Check the printer selection and reselect the standard printer color and/or B&W if needed. Reloading
software may reset the printer selection even after restoring user configurations.
2.) Check that the online help language is set.
3.) Check for any field that appears highlighted in yellow as this is an indication that the field needs to
be reset.
4.) Confirm that at least the system drive encryption has been initialized. Contact the site administrator
for them to set up the user partitions encryption.
Functional checks
Perform the following functional checks to confirm the system is operational before returning the system
to the customer.
Table 8-13 Loading Application Software Functional Checks
: Section
Functional Check
Debrief Script
Equipment passed all required checks and is ready for
use.
Probe/connector functional checks
System CFM and PWD functional checks
Summary of Contents for LOGIQ Totus
Page 2: ......
Page 4: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual ii...
Page 5: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual iii...
Page 6: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual iv...
Page 643: ......