Chapter 4 Functional Checks
4 - 27
4-3-20
Basic measurement functional checks
NOTE:
The following instructions assume that you first scan the patient and then press Freeze.
To check Distance and Tissue Depth Measurement
:
1.) Press
Measure
once to display an active caliper.
2.) Move the trackball to position the active caliper at the start point (distance) or the most anterior point
(tissue depth).
3.) Press
Set
to fix the start point. The LOGIQ Totus fixes the first caliper and displays a second active
caliper.
4.) Move the trackball to position the second active caliper at the end point (distance) or the most
posterior point (tissue depth).
5.) Press
Set
to complete the measurement. The Ultrasound system displays the distance or tissue
depth value in the measurement results window.
NOTE:
Before you complete a measurement:
•
To toggle between active calipers, press
MEASURE
.
•
To erase the second caliper and the current data measured and start the measurement again, press
CLEAR
once.
•
To rotate through and activate previously fixed calipers, turn
Cursor Select
.
After you complete the measurement, to erase all data that has been measured to this point, but not
data entered onto worksheets, press Clear.
To check Circumference/Area (Ellipse) Measurement
:
1.) Press
Measure
once to display an active caliper.
2.) Move the trackball to position the active caliper.
3.) Press
Set
to fix the start point. The LOGIQ Totus fixes the first caliper and displays a second active
caliper.
4.) Move the trackball to position the second caliper.
5.) Turn the
Ellipse
control; an ellipse with an initial circle shape appears.
NOTE:
Be careful not to press the Ellipse control as this activates the Body Pattern.
6.) Move the trackball to position the ellipse and to size the measured axes (the calipers).
7.) To increase the size, turn the Ellipse control in a clockwise direction.
8.) To decrease the size, turn the Ellipse control in a counterclockwise direction.
9.) To toggle between active calipers, press
Measure
.
10.)Press
Set
to complete the measurement. The Ultrasound system displays the circumference and
area in the measurement results window.
Before you complete a measurement:
•
To erase the ellipse and the current data measured, press
Clear
once. The original caliper is
displayed to restart the measurement.
•
To exit the measurement function without completing the measurement, press
Clear
again.
Summary of Contents for LOGIQ Totus
Page 2: ......
Page 4: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual ii...
Page 5: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual iii...
Page 6: ...GE HealthCare CONFIDENTIAL Direction 5936427 Rev 1 LOGIQ Totus Basic Service Manual iv...
Page 643: ......