Chapter 3 LOGIQ Totus Setup
3 - 81
3-7-9-9
Imaging Insights
Use the Imaging Insights page to set parameters for Imaging Insights.
3-7-9-10
Measurement
Use the Measurement page to set parameters for measurements.
Table 3-44 To Configure Imaging Insights
Step
Corresponding Graphic
1. Navigate to
Connectivity > Connection Manager > Advanced Settings
.
2. Select
Imaging Insights
.
3. To store incremental data in each image, enable
Store incremental analytic data in each image
.
4. To send consolidated analytic data to server, enable
Send consolidated analytic data to server
.
5. Under
Image Analytics Server Select
, select the server.
6. Under
Trigger interval outside an exam (hours)
, select Off, 1, 2, 3 or 5.
Table 3-45 To Configure Measurement
Step
Corresponding Graphic
1. Navigate to
Connectivity > Connection Manager > Advanced Settings
.
2. Select
Measurement
.
3. To link measurements to images when sent to PACS, enable
Link Measurement to Images
.
4. To verify that all measurements have image references when sent to PACS, enable
Verify all Measurements have Image References
.